MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/1120700018812993

Медицинская библиотека · Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2018

Adverse reaction to metal debris after small-head diameter metal-on-metal total hip arthroplasty: an increasing concern.

D'Angelo Fabio, Tanas Davide, Gallazzi Enrico, Zagra Luigi

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

: Failures due to adverse reaction to metal debris (ARMD) have become an area of common focus among surgeons performing hip replacements. Several authors have reported data on the prevalence of these masses, in both symptomatic and asymptomatic patients after either large-diameter head metal-on-metal (MoM) total hip arthroplasty (THA) or hip resurfacing arthroplasty, with a large variability of rate. To our knowledge, few data are reported on the association of this lesion with the use of small-head diameter MoM.

METHODS

: 15 hips that were revised for ARMD in small-head MoM THA were included in this study. We focused our attention on the difficulties of diagnosis and treatment and also on the histologic aspects of the harvested pathologic tissue.

RESULTS

: The histological examination of our cases showed a presence of lymphocytic infiltrate suggesting a delayed hypersensitivity reaction to the metal of type IV (ALVAL), but different from each other in term of the prevalence of the cellular component. Osteolysis and severe soft tissue damage were also observed. Revision resulted in remission of the lesion and successful implant.

CONCLUSIONS

: Our observation suggests that the evidence of ARMD should be considered even in case of small-head MoM arthroplasty and therefore these patients should be followed scrupulously with 2nd level diagnostic tools such as magnetic resonance imaging with metal artifact reduction sequence (MARS-MRI) and metal ion levels at least once. Further investigations are necessary to establish the real prevalence of this phenomenon in the whole population of small-head MoM THAs.

Для цитирования: D'Angelo Fabio, Tanas Davide, Gallazzi Enrico, Zagra Luigi. Adverse reaction to metal debris after small-head diameter metal-on-metal total hip arthroplasty: an increasing concern.. Hip international : the journal of clinical and experimental research on hip pathology and therapy. 2018. DOI: 10.1177/1120700018812993.