Maximum temperatures of 89°C recorded during the mechanical preparation of 35 femoral heads for resurfacing.

Baker Richard, Whitehouse Michael, Kilshaw Michael, Pabbruwe Morreica, Spencer Robert, Blom Ashley, Bannister Gordon

СтатьяПеревод в очереди

Аннотация

BACKGROUND AND PURPOSE: We noticed that our instruments were often too hot to touch after preparing the femoral head for resurfacing, and questioned whether the heat generated could exceed temperatures known to cause osteonecrosis.

PATIENTS AND

METHODS

Using an infra-red thermal imaging camera, we measured real-time femoral head temperatures during femoral head reaming in 35 patients undergoing resurfacing hip arthroplasty. 7 patients received an ASR, 8 received a Cormet, and 20 received a Birmingham resurfacing arthroplasty.

RESULTS

The maximum temperature recorded was 89°C. The temperature exceeded 47°C in 28 patients and 70°C in 11. The mean duration of most stages of head preparation was less than 1 min. The mean time exceeded 1 min only on peripheral head reaming of the ASR system. At temperatures lower than 47°C, only 2 femoral heads were exposed long enough to cause osteonecrosis. The highest mean maximum temperatures recorded were 54°C when the proximal femoral head was resected with an oscillating saw and 47°C during peripheral reaming with the crown drill. The modified new Birmingham resurfacing proximal femoral head reamer substantially reduced the maximum temperatures generated. Lavage reduced temperatures to a mean of 18°C.

INTERPRETATION: 11 patients were subjected to temperatures sufficient to cause osteonecrosis secondary to thermal insult, regardless of the duration of reaming. In 2 cases only, the length of reaming was long enough to induce damage at lower temperatures. Lavage and sharp instruments should reduce the risk of thermal insult during hip resurfacing.

Для цитирования: Baker Richard, Whitehouse Michael, Kilshaw Michael, Pabbruwe Morreica, Spencer Robert, Blom Ashley, Bannister Gordon. Maximum temperatures of 89°C recorded during the mechanical preparation of 35 femoral heads for resurfacing.. Acta orthopaedica. 2011. DOI: 10.3109/17453674.2011.636681.