A titanium expandable pedicle screw improves initial pullout strength as compared with standard pedicle screws.

Vishnubhotla Srilakshmi, McGarry William B, Mahar Andrew T, Gelb Daniel E

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Pedicle screws are now standard for spinal arthrodesis as they provide three-column spinal stabilization. Decreased vertebral body bone density because of aging reduces the stability of the bone-screw interface, potentially increasing screw pullout or pseudarthrosis. Modifications to standard pedicle screw designs to improve screw stabilization may help to compensate for the detrimental effects of decreased vertebral bone density.

PURPOSE: To evaluate differences in initial pullout strength of an expandable titanium pedicle screw as compared with a standard titanium pedicle screw.

STUDY

DESIGN

In vitro human cadaveric biomechanical investigation.

METHODS

Fresh thoracolumbar spines from four human cadavers were imaged using quantitative computed tomography to obtain standard lumbar osteoporosis (Dual-energy X-ray absorptiometry [DXA]) T scores. Six bodies were sectioned per spine, and standard titanium 6.5-mm diameter pedicle screws and expandable 6.5-mm diameter titanium screws (maximum expanded diameter=10 mm) were randomized to right and left sides. Screw testing, in axial pullout at 25 mm/min, was randomized to reduce the effects of testing order. Data for stiffness (N/mm), yield load (N), ultimate load (N), and energy (N mm) (area under the load-displacement curve) were analyzed using a one-way analysis of variance (p<.05).

RESULTS

Lumbar DXA scores averaged -3.6. There were no statistical differences between screw types for stiffness. Yield load was not statistically different between groups, although the expandable screw yield load was nearly 25% greater than that of the standard screw. Ultimate load was found to be statistically greater (∼30%) for the expandable screw compared with the standard screw (p<.05). The energy required to cause bone-implant failure was also statistically greater for the expandable screw compared with the standard screw (p<.0001).

CONCLUSIONS

Expandable titanium pedicle screws demonstrated improved screw pullout stability compared with standard titanium screws in osteopenic or osteoporotic bone. Further studies are warranted examining other loading methods to evaluate the stability provided by an expandable pedicle screw.

Для цитирования: Vishnubhotla Srilakshmi, McGarry William B, Mahar Andrew T, Gelb Daniel E. A titanium expandable pedicle screw improves initial pullout strength as compared with standard pedicle screws.. The spine journal : official journal of the North American Spine Society. 2011. DOI: 10.1016/j.spinee.2011.06.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.