DOI 10.1007/s00113-006-1184-4

Медицинская библиотека · Der Unfallchirurg · 2007

[Rotating platform spinout in LCS knee arthroplasty].

Lichtinger T K, Pingsmann A, Saxler G

Клинический случайАннотация на английскомСтатьяПеревод в очереди

Аннотация

In knee arthroplasty, a lax flexion gap may cause flexion instability. Subsequently, the operated knee may subluxate or dislocate. We report on a case of recurrent subluxations with malrotation of the rotating platform (spinout) after primary LCS total knee arthroplasty. Potential causes for spinout, therapy procedures, and consequences for the primary surgical technique are discussed. In the LCS rotating platform knee arthroplasty, the flexion gap must not be laxer than the extension gap. In such an instance, we recommend a higher rotating platform and a more proximal distal transverse femoral osteotomy.

Для цитирования: Lichtinger T K, Pingsmann A, Saxler G. [Rotating platform spinout in LCS knee arthroplasty].. Der Unfallchirurg. 2007. DOI: 10.1007/s00113-006-1184-4.