MedElevator Войти в исследования
← Поиск публикаций

DOI 10.4012/dmj.2024-082

Медицинская библиотека · Dental materials journal · 2024

A new method to measure the mechanical properties of rotary endodontic Ni-Ti files: Measurement of 6-axis force/torque and automatic root canal preparation.

Lee Hyungwoo, Lee In-Bog

СтатьяПеревод в очереди

Аннотация

The purpose of this study was to develop a new instrument to measure the mechanical properties of rotary endodontic Ni-Ti files (ProTaper Gold F2, ProTaper Ultimate F2, and HyFlex EDM Onefile), and to evaluate the overall utility of the device. The instrument was capable of analyzing the 6-axis force/torque generated by the files during cyclic dynamic movement in a metal curved artificial root canal, and doing automatic cyclic dynamic filing in a resin root canal with a preset vertical force limit by adopting a negative feedback mechanism. By analyzing the 6-axis force/torque, we were able to estimate the position and contact points of the files in the curved root canal. ProTaper Gold showed the highest force/torque in all directions. HyFlex EDM had the highest hysteresis ratio, centering ratio value and NCF (number of cycles to fatigue fracture), while the lowest vertical force.

Для цитирования: Lee Hyungwoo, Lee In-Bog. A new method to measure the mechanical properties of rotary endodontic Ni-Ti files: Measurement of 6-axis force/torque and automatic root canal preparation.. Dental materials journal. 2024. DOI: 10.4012/dmj.2024-082.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.