High predicted durability for the novel small-diameter OmniaSecure defibrillation lead.

Crossley George H, Mason Pamela K, Hansky Bert, De Filippo Paolo, Shah Maully J, Philippon Francois, Sholevar Darius, Richardson Travis D, West Michael B, Dinerman Jay, Dawson James, Himes Adam, Severseike Leah, Thompson Amy E, Sanders Prashanthan

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Defibrillation leads remain the Achilles heel of implantable cardioverter-defibrillators. As patients with implantable cardioverter-defibrillators are living longer and battery longevity increases, more durable leads are needed. The LEADR trial evaluated the novel, lumenless, small-diameter, OmniaSecure defibrillation lead and demonstrated favorable safety and efficacy profile as well as zero fractures through 12.7 ± 4.8 months and remains in clinical follow-up. To augment the clinical trial, advanced cardiac lead reliability modeling was used to project long-term lead durability.

OBJECTIVE

We aimed to project the 10-year fracture-free survival of the OmniaSecure defibrillation lead using reliability modeling.

METHODS

The validated reliability model, which incorporates patient and bench test data, was used to project the 10-year fracture-free survival of the OmniaSecure lead. A subset of LEADR trial patients underwent biplane fluoroscopy imaging during cardiac and patient motion to evaluate the lead's bending curvature in vivo. Bench tests then reproduced these use conditions with greater bending curvatures than observed in patients to exaggerate stress on the lead and to evaluate the lead fatigue strength.

RESULTS

The reliability modeling projects a 98.2% fracture-free survival rate of the OmniaSecure lead at 10 years, including a 10-year fracture-free survival rate of 97.9% in adolescents, exceeding both the modeled and clinical 10-year performance of the highly reliable, larger diameter Sprint Quattro lead.

CONCLUSION

Consistent with early clinical trial experience, modeling projects highly durable 10-year performance of the OmniaSecure lead, including within the active adolescent pediatric population, which may uniquely benefit from a novel 4.7F defibrillation lead designed for reliability.

GOV IDENTIFIER: NCT04863664.

Для цитирования: Crossley George H, Mason Pamela K, Hansky Bert, De Filippo Paolo, Shah Maully J, Philippon Francois, Sholevar Darius, Richardson Travis D, West Michael B, Dinerman Jay, Dawson James, Himes Adam, Severseike Leah, Thompson Amy E, Sanders Prashanthan. High predicted durability for the novel small-diameter OmniaSecure defibrillation lead.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.09.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.