[Reliability of porous coating metal-backed cups: advantages and adverse effect of a posterior elevated rim polyethylene liner].

Girard J, Laffargue P, Decoulx J, Migaud H

Аннотация на английскомСтатьяОбзорПеревод в очереди

Аннотация

PURPOSE OF THE STUDY: We analyzed primary implantation of the Duraloc cup associated with a self-locking Muller stem at 9.6 +/- 1.6 years (8.6-11.4). Because of its particular 3/5 semi-spherical shape, the Duraloc cup has been associated with a high rate of dislocation. The objective of this study was to evaluate the efficacy of an anti-dislocation posterior elevated rim polyethylene liner on long-term dislocation rate and wear.

MATERIAL AND

METHODS

The series included 89 arthroplasties implanted between 1991 and 1993 in 82 patients, mean age 56.8 +/- 12.6 years (17.2-87). The Postel-Merle-d'Aubligné (PMA) score and subjective evaluation with a visual analog scale were used for the clinical assessment. Changes in the bone-cup interface, cup migration, and polyethylene wear were assessed radiographically.

RESULTS

At last follow-up, the PMA function score was 16.2 +/- 1.9 points (10-18) (81% good, very good and excellent results). The postoperative x-rays showed a low incidence of lucent lines and osteolysis, respectively 8% and 4%. Mean annual linear polyethylene wear was 0.11 +/- 0.066 mm (0.03-0.57) and only 4% of the cups showed wear greater than 0.2 mm/year. Wear was correlated with the presence of acetabular osteolysis and high activity level. Two cups migrated (3-4 cm medialization). Three cups were revised, one for deep infection, one for recurrent dislocation, and one for dislocation associated with major wear. The rate of dislocation was 2.2% (two cases). Overall survival was 97.3% at 9.6 years (95%CI 0.93-1.0).

DISCUSSION: Compared with other series in the literature using this implant, the presence of a posterior elevated rim polyethylene liner reduced the rate of instability and did not increase wear. The Duraloc cup is recognized as a reliable implant exhibiting excellent osteointegration and a low rate of migration. The presence of a posterior rim is associated with a lower rate of dislocation and does not increase wear. It can thus be proposed for primary implantation. The posterosuperior position for the liner is recommended.

Для цитирования: Girard J, Laffargue P, Decoulx J, Migaud H. [Reliability of porous coating metal-backed cups: advantages and adverse effect of a posterior elevated rim polyethylene liner].. Revue de chirurgie orthopedique et reparatrice de l'appareil moteur. 2005. DOI: 10.1016/s0035-1040(05)84360-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.