Biomechanics of thoracic short versus long fixation after 3-column injury.
Аннотация
OBJECT: Posterior screw-rod fixation for thoracic spine trauma usually involves fusion across long segments. Biomechanical data on screw-based short-segment fixation for thoracic fusion are lacking. The authors compared the effects of spanning short and long segments in the thoracic spine.
METHODS
Seven human spine segments (5 segments from T-2 to T-8; 2 segments from T-3 to T-9) were prepared. Pure-moment loading of 6 Nm was applied to induce flexion, extension, lateral bending, and axial rotation while 3D motion was measured optoelectronically. Normal specimens were tested, and then a wedge fracture was created on the middle vertebra after cutting the posterior ligaments. Five conditions of instrumentation were tested, as follows: Step A, 4-level fixation plus cross-link; Step B, 2-level fixation; Step C, 2-level fixation plus cross-link; Step D, 2-level fixation plus screws at fracture site (index); and Step E, 2-level fixation plus index screws plus cross-link.
RESULTS
Long-segment fixation restricted 2-level range of motion (ROM) during extension and lateral bending significantly better than the most rigid short-segment construct. Adding index screws in short-segment constructs significantly reduced ROM during flexion, lateral bending, and axial rotation (p < 0.03). A cross-link reduced axial rotation ROM (p = 0.001), not affecting other loading directions (p > 0.4).
CONCLUSIONS
Thoracic short-segment fixation provides significantly less stability than long-segment fixation for the injury studied. Adding a cross-link to short fixation improved stability only during axial rotation. Adding a screw at the fracture site improved short-segment stability by an average of 25%.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.