DOI 10.1227/01.neu.0000370200.74098.cc
Comparison of allograft bone and titanium cages for vertebral body replacement in the thoracolumbar spine: a biomechanical study.
Аннотация
BACKGROUND
When an anterior approach to repair a burst fracture is indicated, several devices can be used to restore spinal stability (eg, bone graft, free-standing titanium cage, and expandable titanium cage).
OBJECTIVE
We compare the biomechanical stability and prices of each of these systems.
MATERIALS AND
METHODS
Eight fresh human cadaver T11 through L3 vertebral specimens were harvested and cleaned of soft tissues. T11-T12 and L2-L3 were fixed by screws. The fixed ends were then set in automotive body filler (Bondo). The prepared specimens were tested in the Biaxial Instron tester (8874, Norwood, MA) after a sequence of the following: intact, after the creation of an anterior corpectomy at L1, and after insertion of both of the 2 different titanium cages and the fibular graft. A titanium screw-and-plate anterolateral system was used to secure the construct (VANTAGE, Medtronic Sofamor Danek, Memphis, TN). The conditions of displacement testing were as follows: rotation (+/- 3.5 degrees ), flexion and extension, and left and right bending (+/- 3.5 mm). For each mode of testing, the stiffness was calculated.
RESULTS
The stiffness data, when statistically analyzed by repeated-measures analysis of variance (at P = .05 and power > 0.9), indicated no significant differences among these devices.
CONCLUSION
On the basis of this biomechanical study, the stiffness of the fibular graft was similar to that of the other metallic devices in this cadaver model.