Ergonomic impact of an automated device for endoscopic tool insertion and transfer (with video).

Kim Sang Hyun, Choi Hyuk Soon, Lee Bo Mee, Jeon Han Jo, Lee Jae Min, Kim Eun Sun, Keum Bora, Jeen Yoon Tae, Lee Hong Sik, Kim Bo Ryun, Hwang Joo Ha, Chun Hoon Jai

СтатьяVideo-Audio MediaПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND AND AIMS: Endoscopists frequently experience musculoskeletal injuries, particularly in the distal upper extremities, that are caused by substantial forces, awkward postures, and repetitive movements during procedures. A significant but underexplored risk factor is the repetitive exchange of endoscopic instruments. This study aimed to develop and evaluate an automatic endoscopic instrument insertion and transfer device (INSERTrument; EndoRobotics, Seoul, Republic of Korea) to reduce ergonomic strain and assess its impact on wrist movements and postures.

METHODS

The INSERTrument was evaluated during in vivo gastric endoscopic submucosal dissection (ESD) procedures on porcine models conducted by 2 experienced endoscopists. Outcomes included the number of wrist snaps, total instrument exchange time, and the percentage of time spent in high-risk wrist postures. Inertial measurement units were used to objectively analyze wrist joint angles. The device's performance was compared with conventional manual methods across various endoscopic instruments.

RESULTS

The INSERTrument significantly reduced the number of wrist snaps per instrument exchange (7.3 ± 1.0 vs 69.3 ± 8.3; P < .05) and per ESD procedure (68.6 ± 9.4 vs 656.0 ± 71.8; P < .05), achieving an average reduction of approximately 90% compared with the conventional manual method. The total instrument exchange time per ESD was also significantly reduced in the INSERTrument group compared with the conventional group (127.6 ± 19.4 seconds vs 151.6 ± 10.9 seconds; P < .05). Inertial measurement unit data revealed that the INSERTrument significantly decreased the percentage of time spent in high-risk wrist postures (10.4% ± 2.4% vs 44.4% ± 5.1%; P < .05).

CONCLUSIONS

The use of INSERTrument minimized repetitive wrist movements and high-risk postures associated with endoscopic instrument exchanges. These findings suggest that INSERTrument could improve the ergonomics of endoscopic procedures, potentially reducing the incidence of musculoskeletal injuries among endoscopists. Further studies are warranted to explore the long-term benefits and clinical implications of this device in routine endoscopic practice.

Для цитирования: Kim Sang Hyun, Choi Hyuk Soon, Lee Bo Mee, Jeon Han Jo, Lee Jae Min, Kim Eun Sun, Keum Bora, Jeen Yoon Tae, Lee Hong Sik, Kim Bo Ryun, Hwang Joo Ha, Chun Hoon Jai. Ergonomic impact of an automated device for endoscopic tool insertion and transfer (with video).. Gastrointestinal endoscopy. 2025. DOI: 10.1016/j.gie.2025.02.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.