How much articular displacement can be detected using fluoroscopy for tibial plateau fractures?

Haller Justin M, O'Toole Robert, Graves Matthew, Barei David, Gardner Michael, Kubiak Erik, Nascone Jason, Nork Sean, Presson Angela P, Higgins Thomas F

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

While there is conflicting evidence regarding the importance of anatomic reduction for tibial plateau fractures, there are currently no studies that analyse our ability to grade reduction based on fluoroscopic imaging. The purpose of this study was to determine the accuracy of fluoroscopy in judging tibial plateau articular reduction.

METHODS

Ten embalmed human cadavers were selected. The lateral plateau was sagitally sectioned, and the joint was reduced under direct visualization. Lateral, anterior-posterior (AP), and joint line fluoroscopic views were obtained. The same fluoroscopic views were obtained with 2mm displacement and 5mm displacement. The images were randomised, and eight orthopaedic traumatologists were asked whether the plateau was reduced. Within each pair of conditions (view and displacement from 0mm to 5mm) sensitivity, specificity, and intraclass correlations (ICC) were evaluated.

RESULTS

The AP-lateral view with 5mm displacement yielded the highest accuracy for detecting reduction at 90% (95% CI: 83-94%). For the other conditions, accuracy ranged from (37-83%). Sensitivity was highest for the reduced lateral view (79%, 95% CI: 57-91%). Specificity was highest in the AP-lateral view 98% (95% CI: 93-99%) for 5mm step-off. ICC was perfect for the AP-lateral view with 5mm displacement, but otherwise agreement ranged from poor to moderate at ICC=0.09-0.46. Finally, there was no additional benefit to including the joint-line view with the AP and lateral views.

CONCLUSION

Using both AP and lateral views for 5mm displacement had the highest accuracy, specificity, and ICC. Outside of this scenario, agreement was poor to moderate and accuracy was low. Applying this clinically, direct visualization of the articular surface may be necessary to ensure malreduction less than 5mm.

Для цитирования: Haller Justin M, O'Toole Robert, Graves Matthew, Barei David, Gardner Michael, Kubiak Erik, Nascone Jason, Nork Sean, Presson Angela P, Higgins Thomas F. How much articular displacement can be detected using fluoroscopy for tibial plateau fractures?. Injury. 2015. DOI: 10.1016/j.injury.2015.06.043.