Wear of a highly cross-linked polyethylene liner associated with metallic deposition on a ceramic femoral head.

Della Valle Alejandro González, Doty Stephen, Gradl Guntmar, Labissiere Anthony, Nestor Bryan J

Клинический случайСтатьяПеревод в очереди

Аннотация

A 68-year-old patient underwent revision surgery for recurrent dislocation of a ceramic-on-polyethylene uncemented hip arthroplasty implanted 6 months previously. The aluminum oxide femoral head was found to have titanium deposition measuring 5 by 10 mm in an area articulating with the highly cross-linked polyethylene liner. The liner was observed to possess severe scratching produced by articulation with the titanium deposited on the ceramic head. Despite improved scratch resistance of ceramic-bearing surfaces, irregularities introduced by metallic deposition after recurrent dislocation can damage a highly cross-linked polyethylene liner.

Для цитирования: Della Valle Alejandro González, Doty Stephen, Gradl Guntmar, Labissiere Anthony, Nestor Bryan J. Wear of a highly cross-linked polyethylene liner associated with metallic deposition on a ceramic femoral head.. The Journal of arthroplasty. 2004. DOI: 10.1016/j.arth.2003.12.065.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.