Is stability of femoral neck fractures in the inverted triangle configuration related to the angle between the fracture line and the cannulated compression screws? A finite element analysis.

Niu Zhipeng, Wang Qian, Yuan Baoming, Cui Yutao, Ren Guangkai, Wu Dankai, Peng Chuangang

СтатьяПеревод в очереди

Аннотация

PURPOSE: Currently, treating femoral neck fractures (FNFs) with the inverted triangle configuration requires alignment between the femoral neck's long axis and the axis of cannulated compression screws (CCS). To address whether the 'parallel' alignment is the most effective approach for fractures with varying Pauwels angles, we employed finite element analysis (FEA) to investigate how different angles between fracture line and CCS affect stability, based on various Pauwels angles. This study aims to offer improved guidance for treating FNFs with the inverted triangle configuration.

METHODS

FNF models with Pauwels angles of 40°, 50°, 60°, and 70° were developed. The CCS were positioned in an inverted triangle configuration based on the angle between the fracture line and CCS. Using FEA, we compared the biomechanical properties of each model to evaluate the stability by evaluating five key parameters: maximal stress in the proximal femoral fracture fragment (MPFS) and implants (MIS), maximal displacement of the bone (MBD) and implants (MID), and maximal relative displacement of the fragments (MRD).

RESULTS

For Pauwels angles of 40°, 50°, 60°, and 70° across different FNF models, various parameters exhibited similar results. The MPFS showed an upward trend with a decrease in the angle, whereas the MIS, MBD, MID, and MRD all exhibited downward trends.

CONCLUSION

The FEA results suggest that decreasing the angle between the fracture line and the CCS for the treatment of FNF can increase the tension resistance of the model, thus increasing the model's stability.

Для цитирования: Niu Zhipeng, Wang Qian, Yuan Baoming, Cui Yutao, Ren Guangkai, Wu Dankai, Peng Chuangang. Is stability of femoral neck fractures in the inverted triangle configuration related to the angle between the fracture line and the cannulated compression screws? A finite element analysis.. Computer methods in biomechanics and biomedical engineering. 2024. DOI: 10.1080/10255842.2024.2392556.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.