A cable driven robotic palpation system with contact force sensing based on cable tension observation.
Аннотация
BACKGROUND
Prostate Cancer screening based on manual palpation is subjective. Robotic palpation systems can objectively acquire stiffness conditions of the prostate.
METHODS
A 2DoF cable driven robotic system for prostate palpation is proposed. An indirect method to estimate the contact force based on cable tension observation is presented. Kinematic models and a joint angle estimation method to determine the tip position of the probe are derived. Positioning accuracy was verified using an optical marker tracking system and by displacement measurement, respectively. The contact force estimation method was validated on silicone phantom samples.
RESULTS
A good consistence between the estimated and measured contact force was observed. The contact force was correlated with the elastic modulus of each silicone phantom. There was also a good agreement between the theoretical and the measured tip position.
CONCLUSION
In the proposed palpation system, the indirect contact force estimation method is viable and holds potential for the stiffness assessment of the prostate. The tip position vital for palpation can be determined through estimated joint angles.