Biomechanical stability of a fixed-angle volar plate versus fragment-specific fixation system: cyclic testing in a C2-type distal radius cadaver fracture model.
Аннотация
PURPOSE: To compare the biomechanical stability of 2 recently introduced fixation systems in an intra-articular, dorsal comminution distal radius fracture model.
METHODS
AO/ASIF type C2 fractures were simulated in 10 matched pairs of fresh-frozen cadaveric arms randomized between fixed-angle volar plate and fragment-specific fixation systems. Specimens were loaded in extension cyclically for 2,000 repetitions followed by a single cycle to failure. Initial, intermediate, and final stiffness values and failure load values were obtained and compared.
RESULTS
Both systems were able to sustain physiologic cyclic loading. The fragment-specific system was significantly stiffer than the fixed-angle volar plate system for the ulnar segment in both the precycle and postcycle values. No other comparisons were significant with respect to stiffness. No significant difference in load to failure was found between the systems with respect to ulnar, radial, or overall fragment displacement.
CONCLUSIONS
Both fixed-angle volar plate and fragment-specific fixation systems performed comparably in a simulated early postoperative motion protocol. Fragment-specific fixation had improved stiffness characteristics only with respect to the smaller ulnar-sided fragment.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.