MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3233/thc-240928

Медицинская библиотека · Technology and health care : official journal of the European Society for Engineering and Medicine · 2024

A biomechanical study of a polymer material bundled rib fracture fixator.

Zhang Yongmin, Wang Dongbin, Gong Hao, Tang Haosen, Dong Guangqi, Wang Bin, Xia Honggang

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Rib fractures are one of the most common blunt injuries, accounting for approximately 10% of all trauma patients and 60% of thoracic injuries. Multiple rib fractures, especially flail chest, can cause local chest wall softening due to the loss of rib support, leading to paradoxical breathing, severe pain, and a high likelihood of accompanying lung contusions.

OBJECTIVE

This study investigates the mechanical properties of a new polymer material rib internal fixator to provide theoretical data for its clinical use.

METHODS

We conducted in vitro mechanical tests on 20 fresh caudal fin sheep ribs, using different fracture models across four randomly assigned groups (five ribs per group). The fixators were assessed using non-destructive three-point bending, torsion, and unilateral compression tests, with results averaged. Additionally, finite element analysis compared stress and strain in the polymer fixators and titanium alloy rib plates during bending and torsion tests.

RESULTS

In vitro tests showed that the polymer fixators handled loads effectively up to a maximum without increase beyond a certain displacement. Bending and torsion tests via finite element analysis showed the polymer material sustained lower maximum equivalent stresses (84.455 MPa and 14.426 MPa) compared to titanium alloy plates (219.88 MPa and 46.47 MPa).

CONCLUSION

The polymer rib fixator demonstrated sufficient strength for rib fracture fixation and was superior in stress management compared to titanium alloy plates in both bending and torsion tests, supporting its potential clinical application.

Для цитирования: Zhang Yongmin, Wang Dongbin, Gong Hao, Tang Haosen, Dong Guangqi, Wang Bin, Xia Honggang. A biomechanical study of a polymer material bundled rib fracture fixator.. Technology and health care : official journal of the European Society for Engineering and Medicine. 2024. DOI: 10.3233/thc-240928.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.