Magnetic resonance diagnosis of the anterior labrum and capsule. Effect of field strength on efficacy.
Аннотация
RATIONALE AND
OBJECTIVES
The authors compare the performance of high-field and low-field magnetic resonance (MR) systems in the evaluation of shoulder instability, 35 patients were examined at field strengths of 1.0 T and of 0.2 T.
METHODS
Surface coils were used in both systems. Because arthroscopy was used as gold standard, a preselected patient-population was obtained for the study.
RESULTS
The sensitivity/specificity/accuracy of MR images acquired at 1.0 T for labrum pathology were 91%/67%/91% and 70%/80%/71% for the capsular complex. Compared with the above, the sensitivity/specificity/accuracy for 0.2 T MR images revealed 91%/67%/91% for the labrum pathology and 63%/80%/66% for the capsular complex respectively. In the evaluation of capsular lesions a comparison between the 0.2 T MR system and the 1.0 T system indicated a higher sensitivity and accuracy for the high-field images. Concerning labral lesions, the sensitivity and accuracy of the 0.2 T MR imager and the 1.0 T imager were comparable.
CONCLUSIONS
Given differences in imaging protocols, imaging at 0.2 T does not adversely affect the assessment of shoulder instability when compared with imaging at 1.0 T. These preliminary results warrant more extensive clinical comparison of results obtained at different magnetic field strengths.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.