DOI 10.1016/j.bjps.2024.02.020
BoneTape: A novel osteosynthetic device for the stabilization of zygomatic fractures.
Аннотация
BACKGROUND
The study aims to assess the safety and effectiveness of BoneTape™, a new resorbable bone fixation device, using a zygomatic fracture model in rabbits.
METHODS
The study followed BoneTape™ samples and control (sham) groups over 2-, 6-, and 12-week periods post-zygomaticomaxillary (ZM) osteotomy and zygomaticofrontal (ZF) disarticulation. The osteotomized segments were analyzed for bone healing, inflammatory response, and tissue healing. µCT imaging and histological analysis were used to examine the axial alignment, offset, and quality of new bone formation.
RESULTS
BoneTape™ samples demonstrated enhanced maintenance of the initial intraoperative positioning, reduced axial offset, and better alignment when compared with the control group, enabling stable bone healing under physiological loading conditions. Complete union was observed at 12-weeks in both groups. The BoneTape™ group experienced minimal immune and tissue reactions, classically associated with wound healing, and showed an increased number of giant cells at 6 and 12-weeks.
CONCLUSION
BoneTape™ represents a promising advancement in osteosynthesis, demonstrating efficacy in maintaining stable zygomatic reconstruction and eliciting minimal immune response in a rabbit model. This study introduces BoneTape™ as a disruptive solution specifically designed for clinical application in cranio-maxillofacial fracture fixation, with the potential to eliminate the use of over-engineered solutions while offering benefits such as ease of application and fewer biologically disruptive steps.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.