Locking mechanism failure in the Harris-Galante porous acetabular component associated with recurrent hip dislocation.

Peters Christopher L, Sullivan Charles L

Клинический случайСтатьяОбзорПеревод в очереди

Аннотация

In this review of longitudinal studies of the Harris-Galante porous acetabular component, 20% of all revisions were caused by failure of the polyethylene liner locking mechanism. We report 2 cases of locking mechanism failure in the Harris-Galante porous acetabular component presenting with recurrent dislocation. Broken tines from the acetabular locking mechanism were found embedded into the polyethylene liners in both cases. In a 79-year-old woman with recurrent dislocation, the polyethylene liner and femoral head were replaced, and the acetabular component was revised. In a 74-year-old man with chronic recurrent dislocation, the polyethylene liner and femoral head were replaced, but the acetabular component was preserved. Appropriate treatment for locking mechanism failure with a well-fixed acetabular component depends on many factors, and further long-term data are needed.

Для цитирования: Peters Christopher L, Sullivan Charles L. Locking mechanism failure in the Harris-Galante porous acetabular component associated with recurrent hip dislocation.. The Journal of arthroplasty. 2002. DOI: 10.1054/arth.2002.31075.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.