The Ideal Total Hip Replacement Bearing Surface in the Young Patient: A Prospective Randomized Trial Comparing Alumina Ceramic-On-Ceramic With Ceramic-On-Conventional Polyethylene: 15-Year Follow-Up.

Atrey Amit, Wolfstadt Jesse I, Hussain Nasir, Khoshbin Amir, Ward Sarah, Shahid Mohammad, Schemitsch Emil H, Waddell James P

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

BACKGROUND

The optimum bearing surface for total hip arthroplasty remains debatable. We have previously published our outcome at 10 years and this represents the 15-year follow-up.

METHODS

A total of 58 hips (in 57 patients with a mean age of 42 years) were randomized to receive either ceramic-on-ceramic (CoC) or ceramic-on-polyethylene (CoP) total hip arthroplasty. We prospectively followed for survivorship, functional outcomes (using the Harris Hip Score and the St Michael's Hip Score [SMH]), and radiological outcomes.

RESULTS

At a minimum of 15 years, 3 patients had died, but not been revised. Seven were lost to follow-up. Five cases from the CoP group were revised (4 for polyethylene wear and osteolysis). Four from the CoC were revised; one each for head fracture, instability, infection, and trunnionosis. Both groups showed statistically significant improvements in Harris Hip Score scores and SMH functional scores, with no difference between the 2 bearings. For the CoP group, there was an improvement from 15.6 to 21.5 in the SMH and from 48.8 to 88.7 (P > .05); and for CoC, this improvement was 15.8 to 23.5 and 50.3 to 94.6 (P > .05), respectively. Mean wear rate of the polyethylene was 0.092 mm/y and for the CoC was 0.018 mm/y. Two patients in the CoC group had evidence of acetabular osteolysis vs 3 in the CoP. Six patients had femoral osteolysis in the CoC group and 12 in the CoP group.

CONCLUSION

Survivorship and function of the 2 bearing groups remains comparable; while the polyethylene wear and osteolysis may represent issues in the future.

Для цитирования: Atrey Amit, Wolfstadt Jesse I, Hussain Nasir, Khoshbin Amir, Ward Sarah, Shahid Mohammad, Schemitsch Emil H, Waddell James P. The Ideal Total Hip Replacement Bearing Surface in the Young Patient: A Prospective Randomized Trial Comparing Alumina Ceramic-On-Ceramic With Ceramic-On-Conventional Polyethylene: 15-Year Follow-Up.. The Journal of arthroplasty. 2017. DOI: 10.1016/j.arth.2017.11.066.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.