Proximal tibial fracture stability with intramedullary nail fixation using oblique interlocking screws.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Laflamme G Y, Heimlich D, Stephen D, Kreder H J, Whyne C M
Аннотация
OBJECTIVES: The purpose of this study was to evaluate the mechanical stability of oblique interlocking screws in supplementing intramedullary nail fixation of high proximal tibial fractures.
DESIGN: In vitro experimental testing. SETTING Orthopaedic biomechanics laboratory, Sunnybrook and Women's College Health Sciences Center.
PARTICIPANTS: Ten paired fresh-frozen human cadaver tibiae.
INTERVENTION: One tibia of each pair was randomized to be instrumented with an intramedullary nail (M/DN; Zimmer, Warsaw, Indiana), while the other was stabilized with a 13-hole stainless steel lateral tibial head plate (Synthes AO/ASIF). Specimens were tested in varus-valgus (v/v), flexion-extension (f/e) and torsion, before and after a 2-cm gap osteotomy was performed in the proximal segment. Testing of the nailed tibiae was performed with and without oblique proximal screws. Bone density was physically determined by removing a core of trabecular bone from the distal end of each tibia following testing.
MAIN OUTCOME MEASUREMENT: Biomechanical construct stability.
RESULTS: The addition of the proximally placed oblique screws increased the stability of the nail construct in v/v by 50% (6.8 mm, P < 0.05), in f/e by 47% (7.2 mm, P < 0.05), and in torsion by 18% (3.0 degrees, P < 0.05). There was no significant difference observed between the stability of the intramedullary nail construct with oblique screws and the plated construct. Trabecular bone density had a significant effect in reducing stability (P < 0.05) in nail and plate fixation.
CONCLUSION: The addition of oblique interlocking screws significantly improves the stability of a nailed proximal tibia fracture and provides comparable stability to a plate osteosynthesis.
DESIGN: In vitro experimental testing. SETTING Orthopaedic biomechanics laboratory, Sunnybrook and Women's College Health Sciences Center.
PARTICIPANTS: Ten paired fresh-frozen human cadaver tibiae.
INTERVENTION: One tibia of each pair was randomized to be instrumented with an intramedullary nail (M/DN; Zimmer, Warsaw, Indiana), while the other was stabilized with a 13-hole stainless steel lateral tibial head plate (Synthes AO/ASIF). Specimens were tested in varus-valgus (v/v), flexion-extension (f/e) and torsion, before and after a 2-cm gap osteotomy was performed in the proximal segment. Testing of the nailed tibiae was performed with and without oblique proximal screws. Bone density was physically determined by removing a core of trabecular bone from the distal end of each tibia following testing.
MAIN OUTCOME MEASUREMENT: Biomechanical construct stability.
RESULTS: The addition of the proximally placed oblique screws increased the stability of the nail construct in v/v by 50% (6.8 mm, P < 0.05), in f/e by 47% (7.2 mm, P < 0.05), and in torsion by 18% (3.0 degrees, P < 0.05). There was no significant difference observed between the stability of the intramedullary nail construct with oblique screws and the plated construct. Trabecular bone density had a significant effect in reducing stability (P < 0.05) in nail and plate fixation.
CONCLUSION: The addition of oblique interlocking screws significantly improves the stability of a nailed proximal tibia fracture and provides comparable stability to a plate osteosynthesis.