Torsional stability of cross-link configurations: a biomechanical analysis.

Valdevit Antonio, Kambic Helen E, McLain Robert F

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Cross-link systems have been used to augment segmental spinal instrumentation since the earliest introduction of these fixation systems. Although transverse cross-links have little impact on sagittal motion of spinal constructs, cross-linkage does affect torsional rigidity. Despite the wide variety of cross-link designs, almost all have been configured as transverse devices. The relative mechanical benefit of different cross-link configurations is not known.

PURPOSE: The purpose of this study was to compare the torsional stability of three different cross-link configurations.

STUDY

DESIGN

Biomechanical analysis of segmental instrumentation constructs using porcine spines.

METHODS

Thoracic porcine spines (T4 to T10) were instrumented with 6.5-mm conical pedicle screws and 7.0-mm connecting rods from T5 to T9. Terminal vertebrae were embedded in polymethylmethacrylate (PMMA) after a T7 corpectomy. Four cross-link configurations were tested in a randomized manner: Un-cross-linked Control (CONT); Transverse Rod-Rod (RR); Transverse Screw-Screw (SS); and Diagonal Screw-Screw (DX) Cross-links. The specimens were rotated to 3 Nm at a rate of 0.2 degrees/s and cycled six times with data acquisition over the final two cycles. Stiffness, rotation, and energy data were normalized to each control. A Newman-Keuls repeated measures analysis of variance was used to infer significant differences.

RESULTS

Diagonal cross-link configurations provided the most rigid construct. Transverse cross-links did not significantly change torsional behavior compared with the unlinked control. Rotation and energy expended were not significantly greater torsional stiffness compared with other constructs tested (p<.01).

CONCLUSIONS

The diagonal cross-link configuration provided increased torsional stiffness as compared with unlinked or transverse configurations. This observation should be considered in future cross-link designs.

Для цитирования: Valdevit Antonio, Kambic Helen E, McLain Robert F. Torsional stability of cross-link configurations: a biomechanical analysis.. The spine journal : official journal of the North American Spine Society. 2005. DOI: 10.1016/j.spinee.2005.03.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.