MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.clinph.2012.10.009

Медицинская библиотека · Clinical neurophysiology : official journal of the International Federation of Clinical Neurophysiology · 2012

Non-invasive high voltage electrical stimulation as a monitoring tool of nerve root function in lumbosacral surgery.

Troni Walter, Benech Carlo Alberto, Perez Rossella, Tealdi Stefano, Berardino Maurizio, Benech Franco

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To verify the safety and clinical use of non-invasive high-voltage electrical stimulation (HVES) in patients with compressive radiculopathy. To test the feasibility of HVES to survey nerve root function during lumbosacral surgery.

METHODS

In 20 patients undergoing lumbosacral surgery for degenerative spinal diseases, compound muscle action potentials (CMAPs) evoked by maximal HVES were bilaterally recorded throughout surgery from L3 to S2 radicular territories. A preliminary study was performed in awake patients to rule out detrimental effects caused by HVES.

RESULTS

Preoperative study confirmed the safety of HVES. Unexpectedly, a transient but significant remission of pain was observed after root stimulation. Intraoperative monitoring (IOM) was accomplished in all patients. HVES never hindered surgical procedures and never caused mechanical damage within the operatory field. In 4 patients acute, highly focal and reversible conduction failure was promptly detected by HVES in radicular territories congruent with the root manipulated at that moment.

CONCLUSIONS

HVES is a safe and sensitive tool to monitor nerve root function in lumbosacral surgery.

SIGNIFICANCE: The method is based on the assumption that any acute conduction failure occurring during surgery can be immediately and unambiguously detected by HVES if root stimulation is supramaximal and delivered rostral to the surgical level.

Для цитирования: Troni Walter, Benech Carlo Alberto, Perez Rossella, Tealdi Stefano, Berardino Maurizio, Benech Franco. Non-invasive high voltage electrical stimulation as a monitoring tool of nerve root function in lumbosacral surgery.. Clinical neurophysiology : official journal of the International Federation of Clinical Neurophysiology. 2012. DOI: 10.1016/j.clinph.2012.10.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.