MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2106/jbjs.16.00878

Медицинская библиотека · The Journal of bone and joint surgery. American volume · 2017

Highly Cross-Linked Polyethylene Reduces Wear and Revision Rates in Total Hip Arthroplasty: A 10-Year Double-Blinded Randomized Controlled Trial.

Devane Peter A, Horne James Geoffrey, Ashmore Alexander, Mutimer Jonathon, Kim Woosung, Stanley James

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Highly cross-linked polyethylene (XLPE) was developed to address the problem of wear and osteolysis associated with metal-on-conventional ultra-high molecular weight polyethylene (UHMWPE) bearing surfaces. The purpose of this study was to compare in vivo wear rates and clinical and radiographic outcomes between XLPE and UHMWPE in a prospective double-blinded randomized controlled trial with a minimum of 10 years of follow-up.

METHODS

We randomized 122 patients to receive either a conventional UHMWPE liner (Enduron; DePuy) or an XLPE liner (Marathon; DePuy). Ninety-one patients were assessed clinically and radiographically at a minimum of 10 years (range, 10.08 to 12.17 years). Oxford Hip Scores and Short Form-12 Health Survey scores were collected. The radiographs were analyzed for osteolysis and for 2-dimensional (2-D), 3-dimensional (3-D), and volumetric wear using validated software.

RESULTS

All 122 patients were accounted for at the 10-year follow-up evaluation. Twelve patients had undergone revision surgery, 21 patients had died (1 of whom had previously undergone revision surgery), and 2 patients were unable to return for follow-up, leaving 91 patients available for clinical and radiographic evaluation. At a minimum of 10 years, 3-D wear rates were significantly lower (p < 0.001) in the XLPE group (mean, 0.03 mm/yr) than in the conventional UHMWPE group (mean, 0.27 mm/yr). The prevalence of osteolysis was also significantly lower in the XLPE group (38% versus 8%, p < 0.005), as was the revision rate (14.6% versus 1.9%, p = 0.012), with 10 of the 12 revisions being in the Enduron group. There was no significant difference between the clinical scores of the 2 groups.

CONCLUSIONS

XLPE liners have significantly reduced wear and are associated with a greater implant survival rate at 10 years compared with conventional UHMWPE liners.

LEVEL OF EVIDENCE: Therapeutic Level I. See Instructions for Authors for a complete description of levels of evidence.

Для цитирования: Devane Peter A, Horne James Geoffrey, Ashmore Alexander, Mutimer Jonathon, Kim Woosung, Stanley James. Highly Cross-Linked Polyethylene Reduces Wear and Revision Rates in Total Hip Arthroplasty: A 10-Year Double-Blinded Randomized Controlled Trial.. The Journal of bone and joint surgery. American volume. 2017. DOI: 10.2106/jbjs.16.00878.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.