DOI 10.1016/j.clinbiomech.2021.105516

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2021

Primary stability of cement augmentation in locking plate fixation for proximal humeral fractures: A comparison of absorbable versus non-absorbable cement.

Schöbel Tobias, Schleifenbaum Stefan, Nitsch Viktoria, Hepp Pierre, Theopold Jan

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Cement augmentation has been suggested to increase the stability of screw anchoring in osteoporotic humeral fractures. Initial results are promising but may be jeopardized by cement leakage into the joint and difficult implant removal. Absorbable cement might have advantages in this regard, but it is unclear if the primary stability of both techniques is equivalent to each other. Therefore, this study aimed to compare its primary stability with that of non-absorbable cement augmentation.

METHODS

Nineteen cadaveric humeri with two-part fracture models were treated with locking plate osteosynthesis and cement augmentation using either absorbable calcium phosphate cement (group 1) or polymethylmethacrylate (group 2). Fracture movement, stiffness, failure mode, and ultimate load under cyclic compressive loading were examined and compared between the groups.

FINDINGS: The absolute and relative stiffness values in group 1 were significantly smaller than those in group 2 after 50 cycles (group 1: 114 ± 38 N/mm and 94 ± 8% vs. group 2: 188 ± 71 N/mm and 106 ± 9%; p50 = 0.022), 2000 cycles (group 1: 97 ± 34 N/mm and 81 ± 15% vs. group 2: 153 ± 47 N/mm and 88 ± 15%; p2000 = 0.028), and 5000 cycles (group 1: 98 ± 40 N/mm and 81 ± 22% vs. group 2: 158 ± 40 N/mm and 92 ± 16%; p5000 = 0.028). The failure load was not statistically significantly different between the groups.

INTERPRETATION: Although the PMAA group showed higher values for absolute and relative stiffness, no statistically significant difference was found in the primary stability between absorbable and non-absorbable cement augmentation supporting plate osteosynthesis in proximal humeral fractures. In view of the potential advantages of bio-absorbable cement during the healing process, its use should be considered for the augmentation and stabilization of osteoporotic fractures.

Для цитирования: Schöbel Tobias, Schleifenbaum Stefan, Nitsch Viktoria, Hepp Pierre, Theopold Jan. Primary stability of cement augmentation in locking plate fixation for proximal humeral fractures: A comparison of absorbable versus non-absorbable cement.. Clinical biomechanics (Bristol, Avon). 2021. DOI: 10.1016/j.clinbiomech.2021.105516.