Radiographic evaluation of cementation technique using polished, conical, triple-tapered femoral stem in hip arthroplasty.

Schuroff Ademir Antônio, Deeke Mark, Pedroni Marco Antônio, Lupselo Fernando Silva, Kunz Rodrigo Ernesto, Lima Alexandre Matos

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To radiographically evaluate the quality of cementation and implantation technique using a polished, triple-tapered femoral stem in total hip arthroplasty (THA).

METHOD: Retrospective study with radiographic evaluation of 86 hips in 83 patients who underwent to primary THA with the triple-tapered cemented femoral stem C-Stem (DePuy Orthopedics, Warsaw, Indiana). Cases with at least one-year of follow-up were included, and data related to preoperative, immediate postoperative, and late postoperative radiographic evolution were recorded. This study analyzed, among others, the proximal femoral anatomy, the quality of cementation as described by Barrack, and the implant positioning. Cementation was also evaluated and quantified in the Gruen zones with one-year of follow-up.

RESULTS

The mean age was 62.85 years. Proximal femoral anatomical conformation was Dorr type A in 34 (39.53%) cases, type B in 52 (60.46%), and no type C cases were found. Five (5.81%) cases were defined as type A by Barrack's cementation classification system, 46 (56.49%) type B, 27 (31.40%) type C, and eight (9.30%) type D. The greatest cement mantle thickness was observed in zones four (15.53 mm) and 11 (15.64 mm), and the smallest in zone nine (3.51 mm). Positioning in varus was observed in eight (9.3%) cases, valgus in 25 (29%), forward deviation in two (5%), and backward deviation in 55 (63.95%).

CONCLUSIONS

The C-Stem femoral system presented satisfactory results related to cementation pattern, positioning, osteolysis, and stress shielding with regard to literature referring to double-tapered or triple-tapered models, demonstrating to be a safe method, with a predictable and reliable cementing pattern.

Для цитирования: Schuroff Ademir Antônio, Deeke Mark, Pedroni Marco Antônio, Lupselo Fernando Silva, Kunz Rodrigo Ernesto, Lima Alexandre Matos. Radiographic evaluation of cementation technique using polished, conical, triple-tapered femoral stem in hip arthroplasty.. Revista brasileira de ortopedia. 2017. DOI: 10.1016/j.rboe.2017.08.019.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.