Frondlike Synovitis on MRI and Correlation With Polyethylene Surface Damage of Total Knee Arthroplasty.

Li Angela E, Johnson Christine C, Sneag Darryl B, Koch Chelsea N, Fields Kara, Miller Theodore T, Wright Timothy M, Padgett Douglas E, Potter Hollis G

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The purpose of this study is to determine the association between polyethylene surface damage and the severity of frondlike hypertrophied synovitis, joint distention, capsular thickness, and osteolysis on MRI compared with other patient factors. Another goal of this study is to ascertain the interobserver reliability in MRI evaluation of synovitis.

MATERIALS AND

METHODS

MR images of 61 patients who had undergone revision total knee arthroplasty (TKA) were retrospectively reviewed. The two MRI reviewers were blinded to clinical and retrieval data. The retrieved polyethylene tibial inserts were graded to determine the severity of surface damage. The association of MRI features with surface damage was assessed using Spearman correlation coefficients and multiple linear regression. Interobserver reliability was assessed using the kappa statistic.

RESULTS

Correlations were moderate between the surface damage score and MRI synovitis severity (r = 0.46; p < 0.001) and joint distention (r = 0.41; p = 0.001). Multiple linear regression showed that the MRI synovitis score (p = 0.045), male sex (p = 0.007), body mass index (p = 0.041), and the time since implantation (p < 0.0001) predicted the surface damage score. Substantial interobserver agreement was shown for MRI grading of synovitis (κ = 0.72; 95% CI, 0.65-0.80).

CONCLUSION

The magnitude of frondlike hypertrophied synovitis on MRI is significantly associated with polyethylene surface damage in patients who undergo revision TKA. The finding of severe frondlike synovitis on MR images of patients with TKA should raise the possibility of polyethylene surface damage. However, time since implantation is a stronger predictor of surface damage.

Для цитирования: Li Angela E, Johnson Christine C, Sneag Darryl B, Koch Chelsea N, Fields Kara, Miller Theodore T, Wright Timothy M, Padgett Douglas E, Potter Hollis G. Frondlike Synovitis on MRI and Correlation With Polyethylene Surface Damage of Total Knee Arthroplasty.. AJR. American journal of roentgenology. 2017. DOI: 10.2214/ajr.16.17443.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.