[Biomechanical testing of different ventral fixation devices on the bovine lumbar spine].
Аннотация
AIM: The primary stability of ventral fixation systems has been improved over the past years and special implants supporting minimally invasive procedures have been developed. The aim of this study was to analyze the primary stability of several of these implants.
METHODS
Thirty (30) lumbar segments from 5 to 7-month-old calves were harvested and tested on a pure moment apparatus (PMA) as well as a modified materials testing machine (MTS) according to the European standard recommendations of Wilke et al. Three different implant systems (MACS, Centaur and Z-plate) were included. ROM, neutral zone and stiffness were tested on the PMA up to 10 Nm, stiffness at an eccentric axial load of up to 1500 N was tested on the MTS. Bone density was investigated for all specimens. Statistical analysis was performed by ANOVA and LSD tests.
RESULTS
Primary stability was improved by all fixation devices compared to the physiological specimen. The MACS implant was more stable than the Centaur and Z-plate with the exception of in lateral bending [where the Z-plate was stiffer]. Bone density was comparable in the three different experimental groups.
CONCLUSION
Minimal invasive ventral fixation devices are becoming popular. The primary stability of the minimal invasive MACS was comparable to those of other well-known ventral fixation devices of the lumbar spine.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.