DOI 10.1053/jars.2001.18246

Медицинская библиотека · Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association · 2001

Complications of metallic suture anchors in shoulder surgery: A report of 8 cases.

Kaar T K, Schenck R C, Wirth M A, Rockwood C A

Клинический случайСтатьяПеревод в очереди

Аннотация

PURPOSE: Metallic suture anchors are widely used in open and arthroscopic operations about the shoulder. We report the cases of 8 patients who were referred to our institution with complications following shoulder surgery in which metallic suture anchors were used.

TYPE OF STUDY: Retrospective case series.

METHODS

There were 7 male patients and 1 female patient with an average age of 36 years (range, 18 to 76 years). The initial operation was open anterior reconstruction for anterior instability of the glenohumeral joint in 5 patients, open rotator cuff tear repair in 2 patients, and an open posterior capsular reconstruction for posterior instability in 1 patient. All patients were referred for evaluation after a failed index reconstructive procedure.

RESULTS

On average, 5.5 suture anchors (range, 3 to 8) per shoulder were used. Of 4 patients undergoing reconstruction with glenoid anchors only, 3 patients had an extraosseously positioned device. In this subset of glenoid-sided reconstructions, when more than 3 anchors were used, at least 1 anchor was inserted in an extraosseous position. In 2 of 3 patients with isolated humeral anchors, there was evidence of migration over time (1 intra-articular, 2 bursal). Three patients (38%) developed severe articular damage that was directly caused by a loose or intra-articular metal suture anchor. One patient developed a wound infection after reconstructive surgery. In all 8 patients, the index procedure failed and required subsequent surgery.

CONCLUSIONS

The use of metallic suture anchors about the shoulder is commonplace and useful, but, as with other hardware used about the shoulder, there are significant risks if the anchors are improperly placed or if the index procedure fails.

Для цитирования: Kaar T K, Schenck R C, Wirth M A, Rockwood C A. Complications of metallic suture anchors in shoulder surgery: A report of 8 cases.. Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association. 2001. DOI: 10.1053/jars.2001.18246.