Validation of the suction device Nimble for the assessment of skin fibrosis in systemic sclerosis.

Müller Bettina, Ruby Lisa, Jordan Suzana, Rominger Marga B, Mazza Edoardo, Distler Oliver

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

Skin fibrosis is a main hallmark of systemic sclerosis (SSc). Clinical assessment is done semi-quantitatively using the modified Rodnan skin score (mRSS). Objective measurements for quantifying skin fibrosis could complement the mRSS to achieve higher reproducibility. The aim of this study was to explore the potential of suction measurements to detect structural changes in the skin that are associated with skin fibrosis.

METHODS

This clinical trial included 30 SSc patients and 30 healthy volunteers (HC). We validated a novel suction device-the Nimble-to quantify skin stiffness in comparison to the Cutometer using the OMERACT filter.

RESULTS

A significant difference (p < 0.05) between the skin stiffness of HC and SSc patient groups was found for each location measured. The correlation between the measurements of forearm skin stiffness and the mRSS values was high for the Nimble (r = 0.82) and moderate for the Cutometer (r = 0.58). A ROC analysis showed good ability for the Nimble to distinguish between SSc patients with and without skin involvement (AUC = 0.82). Both suction devices provided excellent reliability in all measurements on HC and SSc patients and proved face validity and feasibility.

CONCLUSION

Suction devices assessing skin stiffness, such as the Nimble, show clear potential to objectively quantify skin fibrosis in SSc patients and might be promising outcome measures complementing established methods such as the mRSS.

TRIAL REGISTRATION: Clinicaltrials.gov, NCT03644225, Registered 23 August 2018-Retrospectively registered, http://www.clinicaltrials.gov.

Для цитирования: Müller Bettina, Ruby Lisa, Jordan Suzana, Rominger Marga B, Mazza Edoardo, Distler Oliver. Validation of the suction device Nimble for the assessment of skin fibrosis in systemic sclerosis.. Arthritis research & therapy. 2020. DOI: 10.1186/s13075-020-02214-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.