Muscle biopsy technical safety and quality using a self-contained, vacuum-assisted biopsy technique.

Gallo Antonio, Abraham Alon, Katzberg Hans D, Ilaalagan Suganthini, Bril Vera, Breiner Ari

СтатьяПеревод в очереди

Аннотация

Muscle sampling via percutaneous biopsy has been shown to be safe and effective using a Bergström needle; however, the use of a single-operator, self-contained, vacuum-assisted biopsy technique has not been explored. We performed a retrospective chart review of muscle biopsy samples obtained using the Vacora® self-contained vacuum-assisted biopsy system between 2013 and 2016, at the Toronto General Hospital. During this period, 102 single-operator muscle biopsies were performed using the Vacora® system. 54/102 showed normal or non-specific findings, while 39/102 were suggestive or diagnostic of a neuromuscular condition. 8 samples did not provide sufficient muscle tissue for diagnosis, and complications (intramuscular hematoma) occurred in 3/102 cases. Mean sample weight was 0.19 grams, which exceeded reported mean sample weights using the suction-modified Bergström technique (0.125 grams). We therefore conclude that the Vacora® vacuum-assisted biopsy system provides a safe and effective method to obtain diagnostic muscle biopsy samples; although complication rates are slightly higher than reported using the Bergström technique, possibly related to more vigorous suction or a sharp needle tip.

Для цитирования: Gallo Antonio, Abraham Alon, Katzberg Hans D, Ilaalagan Suganthini, Bril Vera, Breiner Ari. Muscle biopsy technical safety and quality using a self-contained, vacuum-assisted biopsy technique.. Neuromuscular disorders : NMD. 2018. DOI: 10.1016/j.nmd.2018.02.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.