Novel press-fit technique of patellar bone plug in anterior cruciate ligament reconstruction is comparable to interference screw fixation.

Häberli Janosch, Heilgemeir Maximilian, Valet Sebastian, Aiyangar Ameet, Overes Tom, Henle Philipp, Eggli Stefan

СтатьяПеревод в очереди

Аннотация

PURPOSE: Conventional press-fit technique for anterior cruciate ligament reconstruction (ACLR) is performed with extraction drilling of the femoral bone tunnel and manual shaping of the patellar bone plug. However, the disadvantages of this technique include variation in bone plug size and, thus, the strength of the press-fit fixation, bone loss with debris distribution within the knee joint, potential heat necrosis, and metal wear debris due to abrasion of the guide wire. To overcome these disadvantages, a novel technique involving punching of the femoral bone tunnel and standardized compression of the bone plug was introduced. In this study, the fixation strength and apparent stiffness were tested and compared to that of the gold-standard interference screw fixation technique in three flexion angle configurations (0°/45°/90°) in a porcine model. We hypothesized that the newly developed standardized press fit fixation would not be inferior to the gold standard method.

METHODS

Sixty skeletally mature porcine knees (30 pairs) were used. Full-thickness central third patellar tendon strips were harvested, including a patellar bone cylinder of 9.5 mm in diameter. The specimens were randomly assigned to 10 pairs per loading angle (0°, 45°, 90°). One side of each pair was prepared with the press-fit technique, and the contra-lateral side was prepared with interference screw fixation. Equivalent numbers of left- and right-sided samples were used for both fixation systems. A three-way multifactor ANOVA was carried out to check for the influence of (a) fixation type, (b) flexion angle, and (c) side of the bone pair.

RESULTS

The primary fixation strength of femoral press-fit graft fixation with punched tunnels and standardized bone plug compression did not differ significantly from that of interference screw fixation (p = 0.51), which had mean loads to failure of 422.4 ± 134.6 N and 445.4 ± 135.8 N, respectively. The flexion angle had a significant influence on the maximal load to failure (p = 0.01). Load values were highest in 45° flexion for both fixations. The anatomical side R/L was not a statistically significant factor (p = 0.79).

CONCLUSION

The primary fixation strength of femoral press-fit graft fixation with punched femoral tunnels and standardized bone plug compression is equivalent to that of interference screw fixation in a porcine model. Therefore, the procedure represents an effective method for ACL reconstruction with patellar or quadriceps tendon autografts including a patellar bone plug.

Для цитирования: Häberli Janosch, Heilgemeir Maximilian, Valet Sebastian, Aiyangar Ameet, Overes Tom, Henle Philipp, Eggli Stefan. Novel press-fit technique of patellar bone plug in anterior cruciate ligament reconstruction is comparable to interference screw fixation.. Archives of orthopaedic and trauma surgery. 2021. DOI: 10.1007/s00402-021-04137-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.