DOI 10.1016/j.arth.2007.07.021
Second-generation cementless total hip arthroplasty in patients with osteonecrosis of the femoral head.
Аннотация
The results of cementless total hip arthroplasties in patients with osteonecrosis have been inferior to those in patients with other diagnoses. Fifty-eight primary total hip arthroplasties with insertion of a femoral stem with a circumferentially proximal porous coating and a cementless acetabular component were followed for a mean of 11.1 years. Fifty-seven (98%) of 58 stems were biologically stable, and one stem was loose. There were acetabular revisions in 18 hips (31%) because of polyethylene wear and osteolysis. One hip (1.7%) underwent revision of both acetabular and femoral components. Osteolysis around the acetabular component was seen on radiographs in 22 hips (37.9%). Femoral osteolysis was seen in 9 hips (15.5%), and there was no osteolysis below the lesser trochanter in any hip. Second-generation femoral prostheses provide excellent fixation in patients with osteonecrosis of the femoral head. However, a high rate of polyethylene wear and osteolysis in these high-risk patients remains a challenging problem.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.