Radiographic changes and clinical outcomes associated with an adjustable diaphyseal press-fit humeral stem in primary reverse shoulder arthroplasty.

Harmsen Samuel M, Norris Tom R

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Press-fit humeral fixation in reverse shoulder arthroplasty (RSA) has become increasingly popular; however, radiographic analysis of these stems is limited. We aimed to evaluate the radiographic and clinical outcomes of an adjustable diaphyseal press-fit humeral stem in primary RSA.

METHODS

We conducted a retrospective review of 232 primary RSAs in 219 patients performed by a single surgeon using this system. Radiographic outcomes were evaluated in patients with at least 2 years of radiographic follow-up. Standardized postoperative digital radiographs were analyzed for loosening, osteolysis, and stress shielding. Clinical outcomes in patients who also had complete clinical data sets were evaluated at the most recent follow-up.

RESULTS

Radiographic evidence of loosening was identified in 1 RSA (0.4%) associated with deep infection. Aseptic loosening was not observed. No stems were identified as being at high risk for loosening. Internal stress shielding was observed proximal to the coated diaphyseal component in 226 shoulders (97.4%). This finding was often visible at 3 months (92.7%) and predictably progressed on subsequent radiographs. Progression beyond the 2-year period was rarely seen (4.4%). No external stress shielding or osteolysis was observed. Thirty-six complications occurred in 33 patients (15.1%). At an average follow-up of 36.6 months, significant improvements were identified in all measured clinical outcomes (P < .001).

CONCLUSION

Predictable fixation is achieved using an adjustable diaphyseal press-fit humeral system in primary RSA. Internal stress shielding is commonly observed but does not appear to compromise quality of fixation or clinical outcomes.

Для цитирования: Harmsen Samuel M, Norris Tom R. Radiographic changes and clinical outcomes associated with an adjustable diaphyseal press-fit humeral stem in primary reverse shoulder arthroplasty.. Journal of shoulder and elbow surgery. 2017. DOI: 10.1016/j.jse.2017.02.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.