Analytical and diagnostic performance of an automated anti-CCP assay.

Pavai Sthaneshwar, Sargunan Sokkalingam, Zain Amir Azlan, Chow Sook-Khuan

СтатьяПеревод в очереди

Аннотация

AIM: Autoantibodies against cyclic citrullinated peptide (anti-CCP) are considered to be a sensitive and specific marker for rheumatoid arthritis (RA). This study evaluated the diagnostic and analytical performances of the automated anti-CCP assay.

MATERIALS AND METHOD: Sera from 80 patients with established RA, 65 from other rheumatic diseases (non-RA) and 55 from healthy controls were studied using second generation anti-CCP. Rheumatoid factor (RF) was also assayed in each sample, and the results were compared to the anti-CCP findings. Serum pools were used to determine the precision and linearity.

RESULTS

At a cut-off of 7.4 U/ml for anti-CCP, the sensitivity and specificity for RA were 65% and 98% respectively. RF had a sensitivity of 58% and a lower specificity of 93% than anti-CCP.

CONCLUSION

The high specificity of the assay suggests that anti-CCP is useful in the diagnosis of rheumatoid arthritis and in our cohort of study population anti-CCP exhibits a better diagnostic value than RF. A considerable proportion (28%) of RF-negative RA patients were anti-CCP positive. Based on analytical performance of the assay, we conclude that full automation and high throughput features of AxSYM makes it an ideal platform for routine testing of anti-CCP.

Для цитирования: Pavai Sthaneshwar, Sargunan Sokkalingam, Zain Amir Azlan, Chow Sook-Khuan. Analytical and diagnostic performance of an automated anti-CCP assay.. The Malaysian journal of pathology. 2011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.