BACKGROUND

Mechanically ventilated Intensive Care Unit (ICU) patients often require wrist restraints, contributing to immobility and agitation, over-sedation, and delirium. The Exersides® Refraint® (Healthy Design, LLC), a novel restraint alternative, may be safe and facilitate greater mobility than traditional restraints.

OBJECTIVE

This National Institutes of Health Small Business Technology Transfer (STTR) Program Grant-funded single-site Phase I feasibility study evaluated Exersides® safety and feasibility in anticipation of a multi-site Phase II randomized controlled trial (RCT).

METHODS

In two academic ICUs, mechanically ventilated adults ⩾25 years old who were non-comatose, required restraints and had an expected stay of ⩾2 days were enrolled to wear Exersides® and traditional wrist restraints for 4 h on day 1, in a randomized order, and in the reverse order on day 2. Main outcomes were Exersides® safety (i.e., patient/clinician lacerations/injuries), feasibility (i.e., ⩾90% of required data collected), and patient/family/clinician feedback.

RESULTS

Eight patients were enrolled; one no longer required restraints at initiation, yielding seven subjects (median [interquartile range (IQR)] age 65 [55, 70] years, 86% men). All seven wore Exersides®, averaging (SD) 2.5 (1.0) hours per session, with no safety events reported. Across restraint time periods, 92% and 100% of Richmond Agitation-Sedation Scale (RASS) and wrist actigraphy data, respectively, were collected. Feedback was positive (more movement and comfortable than traditional restraints) and constructive (bulky, intimidating to apply).

CONCLUSIONS

This pilot study provided key safety and feasibility data for a Phase II RCT evaluating Exersides® versus traditional wrist restraints. Feedback motivated minor device modifications before RCT initiation.

Для цитирования: Kamdar Biren B, Fine Janelle M, Pavini Marie T, Ardren Sara S, Burns Stephanie, Bates Jason Ht, McGinnis Ryan S, Pandian Vinciya, Lin Benjamin H, Needham Dale M, Stapleton Renee D. Phase I pilot safety and feasibility of a novel restraint device for critically ill patients requiring mechanical ventilation.. Journal of the Intensive Care Society. 2023. DOI: 10.1177/17511437231182503.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.