MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/00007632-199809150-00005

Медицинская библиотека · Spine · 1998

Biomechanical comparison of C1-C2 posterior fixations. Cable, graft, and screw combinations.

Naderi S, Crawford N R, Song G S, Sonntag V K, Dickman C A

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

STUDY

DESIGN

Four combinations of cable-graft-screw fixation at C1-C2 were compared biomechanically in vitro using nondestructive flexibility testing. Each specimen was instrumented successively using each fixation combination.

OBJECTIVES

To determine the relative amounts of movement at C1-C2 after instrumentation with various combinations of one or two transarticular screws and a posterior cable-secured graft. Also to determine the role of each component of the construct in resisting different types of loading.

SUMMARY OF BACKGROUND DATA: Spinal stiffness increases after instrumentation with two transarticular screws plus a posterior wire-graft compared with a wire-graft alone. Other C1-C2 cable-graft-screw combinations have not been tested.

METHODS

Eight human cadaveric occiput-C3 specimens were loaded nondestructively with pure moments, and nonconstrained motion at C1-C2 was measured. The instrumented states tested were a C1-C2 interposition graft attached with multistranded cable; a cable-graft plus one transarticular screw; two transarticular screws alone; and a cable-graft plus two transarticular screws.

RESULTS

The transarticular screws prevented lateral bending and axial rotation better than the posterior cable-graft. The cable-graft prevented flexion and extension better than the screws. Increasing the number of fixation points often significantly decreased the rotation and translation (paired t test; P < 0.05). Axes of rotation shifted from their normal location toward the hardware.

CONCLUSIONS

It is mechanically advantageous to include as many fixation points as possible when atlantoaxial instability is treated surgically.

Для цитирования: Naderi S, Crawford N R, Song G S, Sonntag V K, Dickman C A. Biomechanical comparison of C1-C2 posterior fixations. Cable, graft, and screw combinations.. Spine. 1998. DOI: 10.1097/00007632-199809150-00005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.