Cross-Reactivity of Chloroquine and Hydroxychloroquine With DRI Amphetamine Immunoassay.

Gomila Isabel, Quesada Loreto, López-Corominas Victoria, Fernández Julia, Servera Miguel Á, Sahuquillo Laura, Dastis Macarena, Torrents Albert, Barceló Bernardino

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Chloroquine and hydroxychloroquine are medical drugs used to treat the chemoprophylaxis of malaria and a second-line anti-inflammatory drug.

METHODS

We performed a study of cross-reactivity of chloroquine and hydroxychloroquine in the DRI Amphetamine Assay inspired by a case report of a self-ingestion of chloroquine after a family dispute, that involved the following: (1) an in vitro study with control samples of healthy subjects, (2) an in vivo study with samples of patients with rheumatoid arthritis, and (3) an evaluation of the cross-reactivity of chloroquine and hydroxychloroquine in 3 additional immunoassays.

RESULTS

In the case report, the Amphetamine DRI assay resulted positive both at 1000 ng/mL cutoff (1507 and 1137 ng/mL) and at 500 ng/mL cutoff (1178 and 642 ng/mL). Chloroquine urine levels were 103,900 and 100,900 ng/mL at 5 and 9 hours after ingestion. The results with control samples showed a positive cross-reactivity of chloroquine in the DRI Amphetamine Assay (approximately 0.74% and 0.89% at cutoff of 1000 and 500 ng/mL, respectively). Hydroxychloroquine did not cross-react with the DRI Amphetamine Assay up to 1,000,000 ng/mL. In patients treated with chloroquine or hydroxychloroquine, DRI Amphetamine did not produce false-positive results. The comparative assay study showed a positive cross-reactivity of chloroquine in the Emit II Plus Amphetamines Assay with control samples.

CONCLUSIONS

Chloroquine can cause false-positive results in the DRI Amphetamine Assay when it is present at high concentrations. Hydroxychloroquine did not produce false-positive results neither in the DRI Amphetamine Assay nor in the others immunoassays evaluated.

Для цитирования: Gomila Isabel, Quesada Loreto, López-Corominas Victoria, Fernández Julia, Servera Miguel Á, Sahuquillo Laura, Dastis Macarena, Torrents Albert, Barceló Bernardino. Cross-Reactivity of Chloroquine and Hydroxychloroquine With DRI Amphetamine Immunoassay.. Therapeutic drug monitoring. 2017. DOI: 10.1097/ftd.0000000000000382.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.