MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1515/cclm.2011.252

Медицинская библиотека · Clinical chemistry and laboratory medicine · 2011

Rapid HLA-B27 screening with real-time TaqMan PCR: a clinical validation in the Dutch population.

Roelandse-Koop Elianne A, Buisman Bert, van Hannen Erik J, van der Zee Anneke, Kortlandt Wouter, Hermans Mirjam H A, van Houte Arend-Jan, van Rhee-Luderer Rianne

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Human leukocyte antigen B27 (HLA-B27) is strongly associated with ankylosing spondylitis. The B27 allele is present in 90% of patients with this disease, whereas it is present in only 9% of Caucasians. Molecular detection of HLA-B27 is traditionally based on allele specific amplification of exon 2 (Olerup method) or exon 3 (Dominguez method) by PCR, followed by gel analysis.

METHODS

We developed a real-time TaqMan PCR based on the Dominguez method with a β-Globin PCR as internal control.

RESULTS

A total of 544 clinical samples were used to compare the real-time TaqMan PCR with the traditional Dominguez PCR, the traditional Olerup PCR and a commercial Olerup based HLA-B27 detection kit (Olerup SSPTM HLA-B27, GenoVision). While 542 samples gave concordant results, two samples showed discrepancies and were further analyzed. One sample that showed a discrepancy was negative with the traditional Olerup method and positive with the three other procedures. Sequencing analysis showed the presence of HLA-B*2712 in this sample. The other sample, positive with both Olerup based PCRs and negative with both Dominguez based methods, turned out to be positive for HLA-B*2707 by sequence analysis.

CONCLUSIONS

With a correct result for 543 out of 544 samples (99.8%), we consider our real-time HLA-B27 PCR is a reliable method to detect HLA-B27 in the Dutch population, with reduced hands-on time and contamination risk compared to traditional PCR methods.

Для цитирования: Roelandse-Koop Elianne A, Buisman Bert, van Hannen Erik J, van der Zee Anneke, Kortlandt Wouter, Hermans Mirjam H A, van Houte Arend-Jan, van Rhee-Luderer Rianne. Rapid HLA-B27 screening with real-time TaqMan PCR: a clinical validation in the Dutch population.. Clinical chemistry and laboratory medicine. 2011. DOI: 10.1515/cclm.2011.252.