Four Cases with Rare Complications of Intramedullary Screw Fixation for Jones Fracture.

Morimoto Yusuke, Komatsu Taichi, Tokuhashi Yasuaki

Клинический случайСтатьяПеревод в очереди

Аннотация

Active treatment with intramedullary screw fixation is now common for athletes with Jones fracture. Outcomes are generally good, but complications can occur. We report 4 rare complications of intramedullary screw fixa-tion. Two cases developed osteomyelitis and pseudarthrosis caused by thermal necrosis. In the other two cases, screw-related complications occurred during the insertion of the tapered headless screw. Although thermal necrosis and screw insertion failures are considered rare complications and not widely reported in the litera-ture, they do occur occasionally. Knowing the mechanisms underlying these complications could help prevent them, and knowing their course could lead caregivers to appropriate interventions when they do occur.

Для цитирования: Morimoto Yusuke, Komatsu Taichi, Tokuhashi Yasuaki. Four Cases with Rare Complications of Intramedullary Screw Fixation for Jones Fracture.. Acta medica Okayama. 2020. DOI: 10.18926/amo/61214.