OBJECTIVE

To summarize the new research progress in distal interlocking screws of cephalomedullary nails for the treatment of intertrochanteric fractures.

METHODS

Relevant domestic and foreign literature was extensively reviewed to summarize the static/dynamic types of distal interlocking screw holes, biomechanical studies, clinical studies and application principles, effects on toggling in the cavity, and related complications of distal interlocking screws.

RESULTS

The mode of the distal interlocking screw holes can be divided into static and dynamic. Distal interlocking screws play the role of anti-rotation, maintaining femur length, resisting compression stress, increasing torque stiffness, resisting varus stress, etc. The number of the screws directly affects the toggling of the main nail in the cavity. At present, regardless of whether long or short nails are used, distal interlocking screws are routinely inserted in clinical practice. However, using distal interlocking screws can significantly increase the duration of anesthesia and operation, increase fluoroscopy exposure time, surgical blood loss, and incision length. There is a trend of trying not to use distal interlocking screws in recent years. No significant difference is found in some studies between the effectiveness of dynamic and static interlocking for AO/Orthopaedic Trauma Association (AO/OTA) 31-A1/2 fractures. At present, the selection of the number and mode of distal interlocking screws is still controversial. When inserting distal interlocking screws, orthopedists should endeavor to minimize the occurrence of complications concerning miss shot, vascular injuries, local stress stimulation, and peri-implant fractures.

CONCLUSION

Distal interlocking screws are mainly used to prevent rotation. For stable fractures with intact lateral walls, long cephalomedullary nails can be used without distal interlocking screws. For any type of intertrochanteric fractures, distal interlocking screws are required when using short cephalomedullary nails for fixation. Different interlocking modes, the number of interlocking screws, and the application prospects of absorbable interlocking screws may be future research directions.

Для цитирования: Qi Yiming, Wang Xiaoxu, Chang Shimin. [Research progress on distal interlocking screws of cephalomedullary nails in intertrochanteric fractures].. Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery. 2024. DOI: 10.7507/1002-1892.202312052.