MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.thromres.2014.09.006

Медицинская библиотека · Thrombosis research · 2014

Evaluation of a chromogenic commercial assay using VWF-73 peptide for ADAMTS13 activity measurement.

Joly Bérangère, Stepanian Alain, Hajage David, Thouzeau Sandrine, Capdenat Sophie, Coppo Paul, Veyradier Agnès

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

INTRODUCTION

Thrombotic thrombocytopenic purpura (TTP) is a thrombotic microangiopathy (TMA), related to a severe functional deficiency of ADAMTS13 activity (<10% of normal). ADAMTS13 activity is thus crucial to confirm the clinical suspicion of TTP, to distinguish it from other TMAs, and to perform the follow-up of TTP patients.

MATERIAL AND

METHODS

We compared the performance of the commercial chromogenic assay Technozym ADAMTS13 Activity ELISA (chromogenic VWF73 substrate, Chr-VWF73, Technoclone, Vienna, Austria), to that of our in-house FRETS-VWF73 used as reference method. A large group of 247 subjects (30 healthy volunteers and 217 patients with miscellaneaous TMAs) was studied.

RESULTS

The lower limit of detection of the Chr-VWF73 was 3%, which is well adapted to the clinically relevant threshold for TTP diagnosis (10%). Our results showed a reasonable agreement between FRETS-VWF73 and Chr-VWF73 assays to distinguish samples with an ADAMTS13 activity <10% from those with an ADAMTS13 activity >10%. However, Chr-VWF73 assay provided false negative results in ~12% of acute TTP patients. Inversely, the Chr-VWF73 assay globally underestimated ADAMTS13 activity in detectable values ranging from 11 to 100% (with a great variability compared to FRETS-VWF73), which may be a concern for the follow-up of TTP patients in remission.

CONCLUSION

In-house assays developed and performed by expert laboratories remain the reference methods that should be used without limitation to control values provided by commercial assays when needed. Also, the development of an international reference preparation will be crucial to improve standardization.

Для цитирования: Joly Bérangère, Stepanian Alain, Hajage David, Thouzeau Sandrine, Capdenat Sophie, Coppo Paul, Veyradier Agnès. Evaluation of a chromogenic commercial assay using VWF-73 peptide for ADAMTS13 activity measurement.. Thrombosis research. 2014. DOI: 10.1016/j.thromres.2014.09.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.