DOI 10.1016/j.jhsa.2020.12.014

Медицинская библиотека · The Journal of hand surgery · 2021

Effect of Ulnar Head Offset on Distal Radioulnar Joint Stability.

Orbay Jorge, Vernon Lauren, Poirier Sophia, Morales Victor, Cambo Richard, Mercer Deana

СтатьяПеревод в очереди

Аннотация

PURPOSE: A complete ulnar head replacement may be indicated in cases of distal radial ulnar joint (DRUJ) dysfunction to address bony pathology in lieu of using a constrained total DRUJ prosthesis. Complete ulnar head implants are simple, but they may be unstable if soft tissue tension is not adequately restored. We hypothesized that incorporating an increased offset in the complete ulnar head replacement would lead to increased tension on the distal oblique interosseous ligament, increased contact force at the DRUJ, and improved joint stability.

METHODS

Using a specially designed jig, we measured instability by comparing displacement under load (stiffness) of the DRUJ in 10 cadaveric specimens under 4 different conditions: (1) intact, (2) native head after excision of the triangular fibrocartilage complex, (3) replacement of the ulnar head with a standard offset ulnar head, and (4) replacement of the ulnar head with an increased offset ulnar head. No soft tissue repair was done. We measured anteroposterior displacement under load with maximum translation of 10 mm or maximum loads of 50 N. We tested all specimens with the forearm positioned in neutral, supination, and pronation.

RESULTS

Excising the triangular fibrocartilage complex decreased the average stiffness of the DRUJ to 46% of the intact state, creating a simulated state of DRUJ instability. Replacing the ulnar head with the standard offset head increased average stiffness to 54% of the intact state. Increasing the ulnar head offset with the simulated total ulnar head replacement increased average stiffness to 77% of the intact state.

CONCLUSIONS

An increased offset ulnar head replacement improves DRUJ stability compared with a standard anatomic offset ulnar head replacement.

CLINICAL RELEVANCE: Understanding DRUJ morphology and offset is important in the treatment of DRUJ arthritis and instability.

Для цитирования: Orbay Jorge, Vernon Lauren, Poirier Sophia, Morales Victor, Cambo Richard, Mercer Deana. Effect of Ulnar Head Offset on Distal Radioulnar Joint Stability.. The Journal of hand surgery. 2021. DOI: 10.1016/j.jhsa.2020.12.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.