Equivalent wear performance of dual mobility bearing compared with standard bearing in total hip arthroplasty: in vitro study.

Gaudin Gaël, Ferreira André, Gaillard Romain, Prudhon Jean Louis, Caton Jacques H, Lustig Sébastien

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: Osteolysis in total hip arthroplasty (THA) depends on polyethylene wear and dictates the survival of the prosthesis. Dual mobility in THAs, which is claimed to reduce dislocation risk, has very good long-term clinical results. However, little is known about how the liner wears in this design, compared to the standard single mobility model.

METHODS

A comparative study looking at wear of a conventional ultra-high-molecular-weight polyethylene liner, using gravimetric measurement, between dual mobility implants and standard implants, was performed on a simulator in accordance with a normed protocol based on the same dimensions, environmental conditions and stresses. A linear regression test was employed.

RESULTS

Under the same conditions (loading, cycles, sterilization, material and surface roughness), the gravimetric wear (for conventional polyethylene) is comparable between a standard and a dual mobility cup. This correlates to ten year follow-up results of dual mobility cup.

DISCUSSION -

CONCLUSION

This in vitro equivalent wear serves to confirm the very good long-term clinical results observed with dual mobility bearing, whose use should not be restricted by concerns about increased polyethylene wear.

Для цитирования: Gaudin Gaël, Ferreira André, Gaillard Romain, Prudhon Jean Louis, Caton Jacques H, Lustig Sébastien. Equivalent wear performance of dual mobility bearing compared with standard bearing in total hip arthroplasty: in vitro study.. International orthopaedics. 2016. DOI: 10.1007/s00264-016-3346-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.