Clinical biomechanics (Bristol, Avon) · 2023
Anterior column acetabulum fracture fixation with a screw-augmented acetabular cup-a biomechanical feasibility study.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Berk Till, Zderic Ivan, Schwarzenberg Peter, Pfeifer Roman, Pastor Tatjana, Halvachizadeh Sascha, Richards R Geoff, Gueorguiev Boyko, Pape Hans-Christoph
Аннотация
BACKROUND: The beneficial effects of unrestricted postoperative full weight bearing for elderly patients suffering hip fractures have been demonstrated. However, there is still existing disagreement regarding acetabular fractures.The aim of this biomechanical study was to evaluate the initial load bearing capabilities of different fixation constructs of anterior column fractures (ACFs) in osteoporotic bone.
METHODS: Artificial pelvises with ACFs were assigned to three groups (n = 8) and fixed with either a 7.3 mm partially threaded antegrade cannulated screw (group AASS), an anteriorly placed 3.5 mm plate (group AAPF), or a press-fit acetabular cup with screw augmentation (group AACF). All specimens underwent ramped loading from 20 N preload to 200 N at a rate of 18 N/s, followed by progressively increasing cyclic testing at 2 Hz until failure performed at a rate of 0.05 N/cycle. Relative displacements of the bone fragments were monitored by motion tracking.
FINDINGS: Initial stiffness (N/mm) was 118.5 ± 34.3 in group AASS, 100.4 ± 57.5 in group AAPF, and 92.9 ± 44.0 in group AACF, with no significant differences between the groups, p = 0.544. Cycles to failure were significantly higher in groups AACF (8364 ± 2243) and AAPF (7827 ± 2881) compared to group AASS (4440 ± 2063), p ≤ 0.041.
INTERPRETATION: From a biomechanical perspective, the minimally invasive cup fixation with screw augmentation demonstrated comparable stability to plate osteosynthesis of ACFs in osteoporotic bone. The results of the present study do not allow to conclusively answer whether immediate full weight bearing following cup fixation shall be allowed. Given its similar performance to plate osteosynthesis, this remains rather an utopic wish and a more conservative approach deems more reasonable.
METHODS: Artificial pelvises with ACFs were assigned to three groups (n = 8) and fixed with either a 7.3 mm partially threaded antegrade cannulated screw (group AASS), an anteriorly placed 3.5 mm plate (group AAPF), or a press-fit acetabular cup with screw augmentation (group AACF). All specimens underwent ramped loading from 20 N preload to 200 N at a rate of 18 N/s, followed by progressively increasing cyclic testing at 2 Hz until failure performed at a rate of 0.05 N/cycle. Relative displacements of the bone fragments were monitored by motion tracking.
FINDINGS: Initial stiffness (N/mm) was 118.5 ± 34.3 in group AASS, 100.4 ± 57.5 in group AAPF, and 92.9 ± 44.0 in group AACF, with no significant differences between the groups, p = 0.544. Cycles to failure were significantly higher in groups AACF (8364 ± 2243) and AAPF (7827 ± 2881) compared to group AASS (4440 ± 2063), p ≤ 0.041.
INTERPRETATION: From a biomechanical perspective, the minimally invasive cup fixation with screw augmentation demonstrated comparable stability to plate osteosynthesis of ACFs in osteoporotic bone. The results of the present study do not allow to conclusively answer whether immediate full weight bearing following cup fixation shall be allowed. Given its similar performance to plate osteosynthesis, this remains rather an utopic wish and a more conservative approach deems more reasonable.