MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1155/2016/7049130

Медицинская библиотека · Case reports in orthopedics · 2016

A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws.

Wichelhaus Alice, Emmerich Judith, Mittlmeier Thomas

СтатьяПеревод в очереди

Аннотация

This article presents a case of implant failure resulting in mechanical instability of a scaphotrapezotrapezoideal arthrodesis using magnesium-based headless bone screws. During revision surgery osteolysis surrounding the screws was observed as well as degraded screw threads already in existence at 6 weeks after implantation. The supposed osseous integration attributed to magnesium-based screws could not be reproduced in this particular case. Thus, it can be reasoned that the use of magnesium-based screws for partial wrist arthrodesis cannot be encouraged, at least not in dual use.

Для цитирования: Wichelhaus Alice, Emmerich Judith, Mittlmeier Thomas. A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws.. Case reports in orthopedics. 2016. DOI: 10.1155/2016/7049130.