MedElevator Войти в исследования
← Поиск публикаций

DOI 10.5301/hipint.5000429

Медицинская библиотека · Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2016

3rd generation alumina-on-alumina in modular hip prosthesis: 13 to 18 years follow-up results.

Toni Aldo, Giardina Federico, Guerra Giovanni, Sudanese Alessandra, Montalti Maurizio, Stea Susanna, Bordini Barbara

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Ceramic-on-ceramic (CoC) couplings are alternative bearings surfaces that have been reported to reduce osteolysis, wear debris and aseptic loosening compared to the use of polyethylene. Early experiences with ceramics had poor results, but they have led to many improvements in the manufacture and design of subsequent implants.

METHODS

We analysed medical files of 300 CoC total hip arthroplasty (THA) with a modular neck performed during period 1995-2000 by a single surgeon for a minimum follow-up of 13 years, evaluating clinical and radiological outcome.

RESULTS

The mean clinical Merle d'Aubigne and Postel hip score at the final follow-up is 17.4, against a preoperative value of 11.4. Overall survivorship with an endpoint of revision is 93.2% (95% CI, 89.0%-97.3%) at 15 years, while considering only prosthesis failures related to aseptic loosening and ceramic breakage, survival rate at 15 years is 97.2% (95% CI, 94.8%-100%). We observed complications that led to revision surgery in 11 patients (4%) (periprosthetic fractures, liner ruptures, septic loosening of the implant, aseptic loosening of the cup, aseptic loosening of the stem). The occurrence of squeaking is low (1.6%, 4 cases) and we analysed the characteristics of these patients.

DISCUSSION: Our study shows an excellent long term survivorship of third generation alumina CoC THA. We reiterate the importance to have a stable implant to maximise the advantage of ceramic and to avoid complications.

Для цитирования: Toni Aldo, Giardina Federico, Guerra Giovanni, Sudanese Alessandra, Montalti Maurizio, Stea Susanna, Bordini Barbara. 3rd generation alumina-on-alumina in modular hip prosthesis: 13 to 18 years follow-up results.. Hip international : the journal of clinical and experimental research on hip pathology and therapy. 2016. DOI: 10.5301/hipint.5000429.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.