OBJECTIVES

Chronic pain spinal implantable electronic devices (CPSIEDs) include devices that provide spinal cord stimulation and intrathecal drug therapy. In this study, we sought to evaluate the trends of CPSIED infections, related complications, and outcomes following the treatment of infection.

MATERIALS AND

METHODS

The Nationwide Inpatient Sample database contains data from 48 states, and the District of Columbia was used to identify patients with a primary diagnosis of CPSIED infection during the years 2005-2014. Patients with intrathecal pumps for the treatment of spasticity were excluded to limit the study population to patients with chronic pain disorders. Treatments were categorized as: 1) without device removal, 2) pulse generator or pump only removal, 3) intrathecal pump system removal, and 4) spinal cord stimulation system removal. Complications associated with CPSIED infections were identified using administrative billing codes.

RESULTS

During the study period 2005-2014, a total of 11,041 patients were admitted to the hospital with CPSIED infections. The majority of the patients were treated without surgical intervention (56%), and a smaller proportion underwent complete system explantation (22.7%). In-hospital mortality or permanent disability due to paralysis after CPSIED infection was around 1.83% and 2.77%, respectively. Infectious complications such as meningitis, abscess formation, and osteomyelitis occurred in 4.93%, 5.08%, and 1.5%, respectively. The median cost of hospitalization was around US $14,118.00, and the median length of stay was approximately six days (interquartile range = 4-13 days).

CONCLUSIONS

The complications of CPSIED infection were higher among patients that did not undergo device removal.

Для цитирования: Goel Vasudha, Kumar Varun, Agrawal Shivani N, Patwardhan Amol M, Ibrahim Mohab, DeSimone Daniel C, Sivanesan Eellan, Banik Ratan K, Shankar Hariharan. Outcomes Associated With Infection of Chronic Pain Spinal Implantable Electronic Devices: Insights From a Nationwide Inpatient Sample Study.. Neuromodulation : journal of the International Neuromodulation Society. 2020. DOI: 10.1111/ner.13263.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.