MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1055/s-0043-1771190

Медицинская библиотека · Veterinary and comparative orthopaedics and traumatology : V.C.O.T · 2023

Use of Locking Plates Fixed with Cortical Screws for Pelvic Fracture Repair in 20 Cats.

Emilian Adrian Dumitru, Genoni Stefano, Morrison Shane, Whitelock Richard

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The aim of this study is to describe the use of String of Pearls (SOP) plates for the surgical management of feline pelvic fractures, including surgical technique, proposed indications, complications, and outcomes in a cohort of cats.

STUDY

DESIGN

This study is a single-center retrospective clinical case series.

METHODS

Medical records (2015-2019) of cats presenting for pelvic fractures (n = 33) were reviewed. Type of fracture, preoperative and postoperative imaging, and complications were retrospectively retrieved. Owners were contacted via questionnaire for medium- and long-term follow-up.

RESULTS

A total of 20 cats met the inclusion criteria. Minor intraoperative complications were encountered in three patients. One patient suffered a major intraoperative complication. Five major complications were encountered postoperatively. These included two greater trochanter osteotomy fixation implant removal and three SOP plate removal. Full function was recovered in all patients according to the owners' assessment.

CLINICAL SIGNIFICANCE: The use of SOP plates in feline pelvic fractures appears to yield a consistently good outcome with a very low incidence of canal narrowing and screw loosening. SOP plates are easy to contour and to apply along the entire pelvic length.

Для цитирования: Emilian Adrian Dumitru, Genoni Stefano, Morrison Shane, Whitelock Richard. Use of Locking Plates Fixed with Cortical Screws for Pelvic Fracture Repair in 20 Cats.. Veterinary and comparative orthopaedics and traumatology : V.C.O.T. 2023. DOI: 10.1055/s-0043-1771190.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.