Feasibility and Accuracy of Wrist-Worn Sensors for Perioperative Monitoring During and After Major Abdominal Surgery: An Observational Study.

Xu William, Wells Cameron I, Seo Sean Hb, Sebaratnam Gabrielle, Calder Stefan, Gharibans Armen, Bissett Ian P, O'Grady Gregory

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Continuous, ambulatory perioperative monitoring using wearable devices has shown promise for earlier detection of physiological deterioration and postoperative complications, preventing 'failure-to-rescue'. This study aimed to compare the accuracy of vital signs measured by wrist-based wearables with gold standard measurements from vital signs monitors or nurse assessments in major abdominal surgery.

METHODS

Adult patients were eligible for inclusion in this prospective observational study validating the Empatica E4 wrist sensor intraoperatively and postoperatively. The primary outcomes were the 95% limits of agreement (LoA) between manual and device recordings of heart rate (HR) and temperature evaluated via Bland-Altman analysis. Secondary analysis was conducted using Clarke-Error grid analysis.

RESULTS

Overall, 31 patients were recruited, and 27 patients completed the study. The median duration of recording per patient was 70.3 h, and a total of 2112 h of data recording were completed. Wrist-based HR measurement was accurate and moderately precise (bias: 0.3 bpm; 95% LoA -15.5 to 17.1), but temperature measurement was neither accurate nor precise (bias -2.2°C; 95% LoA -6.0 to 1.6). On Clarke-Error grid analysis, 74.5% and 29.6% of HR and temperature measurements, respectively, fell within the acceptable range of reference standards.

CONCLUSIONS

Continuous perioperative monitoring of HR and temperature after major abdominal surgery using wrist-based sensors is feasible but was limited in this study by low precision. While wrist-based devices offer promise for the continuous monitoring of high-risk surgical patients, current technology is inadequate. Ongoing device hardware and software innovation with robust validation is required before such technologies can be routinely adopted in clinical practice.

Для цитирования: Xu William, Wells Cameron I, Seo Sean Hb, Sebaratnam Gabrielle, Calder Stefan, Gharibans Armen, Bissett Ian P, O'Grady Gregory. Feasibility and Accuracy of Wrist-Worn Sensors for Perioperative Monitoring During and After Major Abdominal Surgery: An Observational Study.. The Journal of surgical research. 2024. DOI: 10.1016/j.jss.2024.06.038.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.